Análisis del ambiente electromagnético en un área de estomatología

Analysis of the Electromagnetic Environment in a Dentistry Area

Authors

  • Alejandro Rivas Gamallo Universidad Tecnologica de La Habana “Jose Antonio Echeverria”, CUJAE. La Habana
  • Yenisel Díaz Roller Universidad Tecnologica de La Habana “Jose Antonio Echeverria”, CUJAE. La Habana
  • Janet Carvajal de la Osa Universidad Tecnologica de La Habana “Jose Antonio Echeverria”, CUJAE. La Habana
  • Rut Lay Abad Rodríguez Universidad Nacional Autónoma de México(UNAM). Ciudad de México( D.F.)
  • Joel David Gago García Universidad Nacional Autónoma de México(UNAM). Ciudad de México( D.F.)

Abstract

Dental consultations are highly frequented in the healt area. Several medical devices converge in these settings. The use of electrical energy generates electromagnetic fields that affect the environment. To control this, there are recommendations issued by the International Commission on Non-Ionizing Radiation Protection. Therefore, measurements of the existing magnetic field in the dental consultation area of the CUJAE hospital were conducted to verify if the environmental values are within permitted limits using an EMDEX gaussmeter. The results indicated that when the equipment is operational, values exceeding 120 mG are reached near the pedestal. Although these levels are below the recommended limits, it should be borne in mind the levels they reach, especially since there is only one dental chair located in this area.

Published

2024-12-19